Printout Information
For troubleshooting, or even daily log information, the "detail" report format is preferable. The Detail Report provides a
complete description of the sensors for a calibration or sample. Information for all the sensors, as well as the temperature
probe, is included.
11.1.1 Enzyme Sensors
Listed below are example printouts and explanations of the Detail format information for enzyme sensors.
Sample Report (Detail) Calibration Report (Detail)
======================= =======================
Sample Results Report Cal Report [1]
======================= =======================
Batch: Test Batch-1 Cal Shift
1A:Glucose 2.50 g/L
Analyte P96_A01 IB nA 1.79
----------------------- NPL nA 17.63
1A:Glucose 4.82 g/L FB nA 1.51
IB nA 2.11 PL Slope nA/m 0.42
NPL nA 32.65 IB Shift -0.65%
PL Slope nA/m 0.69 NPL Shift 3.64%
Temp (C) 25.9 Temp (C) 25.9
----------------------- -----------------------
Volume (uL) 25 1B:Lactate 0.50 g/L
Dilution Factor x1 IB nA 0.88
Fri 9/9/2020 10:59:37 NPL nA 9.81
YSI 2950 – 20F000025 FB nA 0.76
======================= PL Slope nA/m 0.42
IB Shift -1.95%
NPL Shift 1.64%
======================= Temp (C) 25.9
Sample Results Report -----------------------
======================= End Point (sec) 30
Batch: Test Batch-1 Volume (uL) 25
Analyte: P96_A01 Fri 9/9/2020 10:59:37
----------------------- YSI 2950 – 20F000025
1B:Lactate 1.82 g/L =======================
IB nA 2.11
NPL nA 32.65
PL Slope nA/m 0.69
Temp (C) 25.9
-----------------------
Volume (uL) 25
Dilution Factor x1
Fri 9/9/2020 10:59:37
YSI 2950 – 20F000025
=======================