10-1
10. Storage
During normal use, the 2900 Series should be left with the power on at all times. It should also have an adequate supply
of buffer for any installed Membranes or ISEs. This will keep the enzyme probes polarized and ready for use and prevent
the reference ISE from drying out.
Instrument Storage
If the 2900 Series is not going to be used for 2 weeks or longer, remove the buffer and calibrator solutions from the bottles
and replace them with authorized rinsing agent (see Section 9.1). Flush the rinsing agent through the system thoroughly,
empty the bottles, reinstall them in the instrument and prime the system with air. After all the fluid has been pumped from
the system, drain the sample modules by temporarily removing one probe. Reinstall the enzyme probes and store the
instrument with the membranes in place.
Store the instrument in an environment from 1535°C, 1075% humidity (non-condensing).
Enzyme Membrane Storage
Extra membranes should be refrigerated until use. Once installed, Membranes should remain in the appropriate buffer
solution and not allowed to dry out.
ISE Storage
10.3.1 Reference ISE
Do not allow the reference ISE to dry out. Always store it wet (in ISE buffer solution).
10.3.2 Ammonium/Potassium ISE
For storage of 5 days or less, store the ammonium and potassium ISEs wet (in ISE buffer solution). For storage greater
than 5 days, store the ammonium and potassium ISEs dry and in a dark enclosure.
Instrument Handling/Transport
Before transporting, drain all fluids as described above and secure the sipper assembly.
Transporting the instrument may require two people.