Sipper pump head
Figure 9-4
Pull the white pump base from the clear pump housing. Immerse the clear pump housing in authorized disinfecting agent
(see Section 9.1). After cleaning, rinse the pump housing for 5 minutes with authorized rinsing agent (see Section 9.1).
Wipe dry with a lint-free tissue. Make sure the metal pipes where the tubing connects are not blocked or restricted.
Replace the O-ring seals as shown in Figure 9-5. New seals are supplied in the Preventive Maintenance Kit. Be sure to
reinstall the black spacer between the two small red O-rings.
Sipper pump seal replacement
Figure 9-5
Reassemble the pump, position the plunger as shown in Figure 9-6 and install it back on the instrument.
Sipper pump plunger position
SPACER
RED O-RING
RED O-RING