YSI 2900D handleiding

172 pagina's
PDF beschikbaar

Handleiding

Je bekijkt pagina 130 van 172
9-6
Sipper pump head
Figure 9-4
Pull the white pump base from the clear pump housing. Immerse the clear pump housing in authorized disinfecting agent
(see Section 9.1). After cleaning, rinse the pump housing for 5 minutes with authorized rinsing agent (see Section 9.1).
Wipe dry with a lint-free tissue. Make sure the metal pipes where the tubing connects are not blocked or restricted.
Replace the O-ring seals as shown in Figure 9-5. New seals are supplied in the Preventive Maintenance Kit. Be sure to
reinstall the black spacer between the two small red O-rings.
Sipper pump seal replacement
Figure 9-5
Reassemble the pump, position the plunger as shown in Figure 9-6 and install it back on the instrument.
Sipper pump plunger position
SPACER
RED O-RING
RED O-RING

Bekijk gratis de handleiding van YSI 2900D, stel vragen en lees de antwoorden op veelvoorkomende problemen, of gebruik onze assistent om sneller informatie in de handleiding te vinden of uitleg te krijgen over specifieke functies.

Productinformatie

MerkYSI
Model2900D
CategorieNiet gecategoriseerd
TaalNederlands
Grootte50780 MB