9-4
Figure 9-1
Unscrew and remove the enzyme/ISE probes and the temperature probe from each sample module. Place the ISEs and
reference electrodes in 2970 Buffer solution to prevent them from drying out.
Figure 9-2
Remove the sample modules. Remove and discard the small magnetic stir bar inside each module. Immerse the modules
in the authorized disinfecting agent (see Section 9.1) for a maximum of 10 minutes. If soiling or residue is visible, the
module may be immersed in a water filled room temperature sonification bath for a maximum of 10 minutes. After
cleaning, rinse the modules for 3 to 5 minutes with authorized rinsing agent (see Section 9.1). Wipe dry with a lint-free
tissue.
9.4.2 Waste Module Cleaning
Unscrew the three hex screws from the top of waste module 1. Disconnect the calibrator and waste tubing and remove
the waste modules from the base plate.
Thumbnuts
Sample module
Enzyme Probe
Enzyme Probe
Potassium ISE
Temperature Probe
Reference Electrode
Ammonium ISE