User manual - MA964802 19  
8.5  Acquisition of potentiometer limits
This function allows you to activate an "on eld" calibration procedure of the 
minimum and maximum limits for the potentiometer input.
This function is enabled for conguration in the “Potentiom. input” section at 
the item “Acquisition”. Once acquisition has been enabled and you have exited 
confi uration, the following screen appears automatically.
Min
Acquired value
Max
0
100
Esc
48
At this point, proceed as follows:
• Set the potentiometer at the position you wish to associate to the 
Minimum 
value
.
• Press
 to acquire the position and to store the value. The 
Acquired value
will indicate the 
Min.
 value showing that the command has been executed.
• Set the potentiometer at the position you wish to associate to the 
Maximum 
value
.
• Press
 to acquire the position and to store the value. The 
Acquired value
will indicate the 
Max.
 value showing that the command has been executed.
At this point, the potentiometer acquisition procedure is nished; check that the 
potentiometer position and the 
Acquired value
 correspond. 
Press 
 Esc
 to exit the procedure.