The green earthing potential socket (3) is connected to the earthing conductor directly. The
adjustable output can be used with the positive terminal to ground reference, or the nega-
tive terminal to ground reference or oating.
Enable the output on the power pack by pressing the „OUTPUT ON / OFF“ (2) button.
Switch the consumer on. The current indicator “A” displays the current that the connected
load draws.
USB loads
The output voltage and current limitations set via the adjustment knobs do not apply to
the USB ports “AUX.1” and “AUX.2”. Voltage and current for the USB ports are controlled
automatically. Refer to the “Technical data” for voltage and current supply limits.
You can check the current output to the USB sockets (8/9) via the indicator “AUX.1” for
USB output “AUX.1” and the indicator “AUX.2” for output “AUX.2”. Current is displayed in
Ampere “A”.
Check the currently supplied voltage on USB socket “AUX.1” by pressing and holding the
adjustment control “VOLTAGE” for 3 seconds and then releasing it.
– The display shows the supplied voltage for 3 seconds and then shows the rated volt-
age again.
Check the voltage supplied on USB socket “AUX.2” by pressing and holding the adjustment
control “CURRENT” for 3 seconds and then releasing it.
– The display shows the supplied voltage for 3 seconds and then shows the rated volt-
age again.
10 Remote control
The laboratory power supply can be controlled from a Windows
®
computer via USB connection.
You will need to install a driver and software.
Driver, software and software operating instructions are available from www.conrad.com/down-
loads.
10.1 Installing driver and software
Download the driver, software from www.conrad.com/downloads.
Install the driver rst then install the software.
10.2 Connecting computer and laboratory power supply
Connect the supplied USB cable to the USB-B port “Remote control” at the rear of the
laboratory power supply.
Connect the USB-A plug of the USB cable to the computer.
– The remote control symbol indicates an active remote connection.
10.3 Remote controlling the laboratory power supply
Download the software operating instructions from www.conrad.com/downloads.
Remotely control the laboratory power supply according to the software operating instruc-
tions.
11 Maintenance and Cleaning
Always unplug the power plug from the mains socket before cleaning the
device.
Do not use aggressive cleaning agents, rubbing alcohol or other chemical solutions. They
damage the housing and can cause the product to malfunction.
Do not immerse the product in water.
Except for replacing the fuse and occasional cleaning, the power supply unit does not
require any maintenance.
Use a dry, clean, lint-free, antistatic cloth to clean the device. Do not use any abrasive or
chemical agents or detergents containing solvents.
12 Replacing the fuse
If it is no longer possible to switch on the power unit, the mains fuse is probably defective.
Turn off the laboratory power supply.
Remove all connecting cables. Pull the mains plug from the rear
connection (11).
Remove the main fuse holder (10) with a suitable screwdriver from
the bracket, as displayed.
Replace the defective fuse with a new one of the same type. Refer
to the “Technical data” for the fuse specications.
Push the main fuse holder carefully into the fuse holder.
13 Troubleshooting
Always observe the safety instructions!
Problem/event Solution/explanation
Error code F01 The overvoltage protection is activated.
Switch the laboratory power supply off.
Disconnect all connected consumers.
Switch the laboratory power supply on again.
Error code F02 The overcurrent protection is activated.
Disconnect all connected consumers.
Switch the laboratory power supply off.
Switch the laboratory power supply on again.
Error code F03 The overheat protection is activated.
Disconnect all connected consumers.
Switch the laboratory power supply off.
Let the laboratory power supply cool down for at least 30
minutes.
Check if air ventilation slots at the side, the rear and the
bottom are clean and free of dust.
The display is not illumi-
nated.
No mains voltage?
Check the safety switch for the mains socket
Check whether the mains plug is seated correctly in the
mains socket
Check the mains fuse in the power supply unit.
Connected loads don't
work.
Correct voltage set?
Is the polarity correct?
Has the power supply overloaded or is the current limitation
(display "CC") active?
Check the technical data of the load.
Output switched off manually?
14 Disposal
This symbol must appear on any electrical and electronic equipment placed on the
EU market. This symbol indicates that this device should not be disposed of as
unsorted municipal waste at the end of its service life.
Owners of WEEE (Waste from Electrical and Electronic Equipment) shall dispose
of it separately from unsorted municipal waste. Spent batteries and accumulators,
which are not enclosed by the WEEE, as well as lamps that can be removed from
the WEEE in a non-destructive manner, must be removed by end users from the
WEEE in a non-destructive manner before it is handed over to a collection point.
Distributors of electrical and electronic equipment are legally obliged to provide free take-back
of waste. Conrad provides the following return options free of charge (more details on our
website):
in our Conrad ofces
at the Conrad collection points
at the collection points of public waste management authorities or the collection points set
up by manufacturers or distributors within the meaning of the ElektroG
End users are responsible for deleting personal data from the WEEE to be disposed of.
It should be noted that different obligations about the return or recycling of WEEE may apply in
countries outside of Germany.
Bekijk gratis de handleiding van Voltcraft TOPS-3363N, stel vragen en lees de antwoorden op veelvoorkomende problemen, of gebruik onze assistent om sneller informatie in de handleiding te vinden of uitleg te krijgen over specifieke functies.
Productinformatie
| Merk | Voltcraft |
| Model | TOPS-3363N |
| Categorie | Niet gecategoriseerd |
| Taal | Nederlands |
| Grootte | 2954 MB |