Matters needing attention
1Clean the glass before it is installed.
2 The standard mouth and the sealing surface should be coated with a
small amount of vacuum grease. Increase the air tightness.
3The installation should lock the head position and cannot move.
METHOD OF USE
The evaporation rate of the instrument depends on three terms.
1. The higher the vacuum degree is, the higher the speed of the
evaporation, the vacuum degree should not be less than that of the water
evaporation 0.096Mpa.
2. The higher the temperature of the water bath, the faster it is.Do not
generally exceed 80, specific process requirements.
3. The lower the cooling water temperature is, the faster the cooling water
is, and the effect of cooling water is better.
An operating program:
1. After vacuuming and opening the vacuum pump, it is found that the
vacuum is not able to be enough . check whether the bottle ports are
sealed well, whether the vacuum pumps leak or not, whether the sealing
ring of the rotating shaft is fixed, and that a vacuum switch connected to
the vacuum tube can be used to fine tune the best recovery rate and
evaporation speed.
2. Feed: using the vacuum pressure of the system, the liquid can be
inhaled with a hose in the feeding port, and the liquid should not exceed
half of the rotating bottle. The instrument can be continuous feeding,
should pay attention to when feeding off the vacuum pump. The stop after
stop heating to evaporate slowly open the feeding pipe to prevent back
flow of cocks.
3. Heating: This instrument is equipped with a specially designed water
bath pot. It