TensCare iTouch Go handleiding
Handleiding
Je bekijkt pagina 9 van 55

16 17
No
sensation
and
display is
ashing
The unit is faulty You can try to test the unit by
holding the probe in your hand:
Dampen your hand with water
and a little table salt. Squeeze
theprobermlyandmakesure
yourskiniscoveringthemetal
parts of the probe and carefully
increase strength until you can feel
something. Most people will start to
feel the stimulation in their hand at
around 25 mA.
Ifthedisplayashesandtheunit
will not allow you to pass 20 mA.
The unit is faulty.
Intensity
setting
returns to
0mA.
There is an open
circuit leading
to a connection
error.
The iTouch Go wireless probe is
unabletoprovidetreatmentdue
to poor/no connection. Ensure the
positioning of the probe is correct.
No
sensation
on one
side of the
probe
Position is not
optimal and
needs adjusting.
Thecurrentowsfromoneside
of the probe to the other, so it is
notpossibletohaveoneside“not
working”.However,thestrength
of the sensation depends on how
closetothenervethecurrentows,
andinwhichdirectionitows
relativetothenerve.Youcantry
slightly adjusting the position of
the probe.
Sudden
change in
sensation
If you disconnect
and re-connect
a few minutes
later, the signal
will feel quite a
lot stronger.
Always return strength to zero after
disconnecting the probe.
Iftheabovereviewhasfailedto
resolveyourproblem,toreport
unexpectedoperationorevents,
ortoprovidefeedbackcall
TensCare or your local dealer
(addressonbackcover)for
advice.
Contact TensCare customer
serviceon+44(0)1372723434.
Our staff are trained to assist you
withmostissuesyoumayhave
experienced, without the need to
sendyourproductinforservice.
EuropeanMedicalDevice
Regulation requires that any
serious incident that has
occurred in relation to this
deviceshouldbereported
to the manufacturer and the
competent authority in your
country. This can be found at:
https://ec.europa.eu/
docsroom/documents/36683/
attachments/1/translations/en/
renditions/pdf
13. EMC
Wireless communications
equipment such as wireless
homenetworkdevices,mobile
phones, cordless telephones and
their base stations, walkie-talkies
can affect this equipment and
should be used no closer than
30cm (12 inches) to any part of
thedevice.
(Note. As indicated in 5.2.1.1(f) of
IEC 60601-1-2 for ME EQUIPMENT).
Note: See end of user
manual for full EMC tables.
14. Disposal of
electronic
products (WEEE)
Oneoftheprovisionsofthe
EuropeanDirective2012/19/
EU is that anything electrical
or electronic should not be
treated as domestic waste and
simply thrown away. To remind
youofthisDirectiveallaffected
products are now being marked
with a crossed-out wheelie bin
symbol, as depicted below.
TocomplywiththeDirective,you
can return your old electro-
therapy unit to us for disposal.
Simply print a postage- paid
PACKETPOST RETURNS label from
our website www.tenscare.
co.uk,attachthistoanenvelope
or padded bag with the unit
enclosed, and post it back to
us. Upon receipt, we will process
yourolddeviceforcomponents
recoveryandrecyclingtohelp
conservetheworld’sresources
andminimiseadverseeffectson
theenvironment.
15. Technical specication
Waveform Symmetrical Bi-Phasic Rectangular
Amplitude
(over 500 Ohm load)
99.0 mA zero to peak (99 steps of
intensity) +/- 10%
Maximum intensity 50Vzerotopeakacross500Ω+/-10%
Constantvoltageover500-2000Ohm
Maximum charge per
pulse
25uC
Output plug Fully shielded: touch proof
Channels Single channel
Batteries Controller: 3.7V/900mAh
Probe:3.7V/320mAh
Weight Wireless probe: 54g
Control hub: 173g
Total: 227g
Dimensions Wireless probe: 26 x 157 x 26 mm
Control hub: 173 x 81.5 x 40 mm