1. Intended use
iTouchGoisamedicaldevice
designed to be used in the home
healthcareenvironmenttotreat
symptoms of urinary and/or
faecal incontinence.
iTouchGoalsohelpstorelieve
symptomsofchronicpelvicpain.
iTouch Go is suitable for use
by adults (18 years+) who
cancontrolthedeviceand
understand the instructions.
Donotusethedeviceforany
purpose other than this intended
use.
2. Kit Contents
1 x iTouch Go Wireless probe
1 x iTouch Go Control hub
1 x Go Gel (100ml)
1 x Type-C charging cable
1 x Instructions for use
3. Safety
guidelines
Contraindications
Do NOT use if you are or may
be pregnant. It is not known
whether TENS may affect foetal
development.
DoNOTuseifyouhavea
pacemaker(orifyouhavea
heart rhythm problem) or with
anyelectronicmedicaldevices.
Using this unit with electronic
medicaldevicesmaycause
erroneous operation of the
device.Stimulationinthedirect
vicinityofanimplanteddevice
may affect some models.
DoNOTuseintherst6weeks
followingapelvicsurgeryor
vaginalchildbirth.Stimulation
may disrupt the healing process.
DoNOTuseifyouhave
symptomsofactiveurinarytract
infection,vaginalinfections,or
localized lesions. Introducing
theprobemayirritatesensitive
tissue.
DoNOTuseifyouhavepoor
sensationinthepelvicregion.
You may not be able to control
the intensity of stimulation safely.
Do NOT use with electrical
stimulation in patients with a
history of carcinoma at the site of
stimulation.
Warnings
Do NOT use if you are unable to
properlyinsertthevaginalprobe.
Ifyouhaveasevereprolapse,or
if any discomfort occurs when
inserting the probe, consult your
healthcare professional before
use.
DoNOTusewhendriving,
operating machinery, or similar
actions needing muscular
control. Damaged leads or
sudden changes in contact may
causebriefinvoluntarymuscle
movements.
DoNOTusetomaskorrelieve
undiagnosed pain. This may
delaydiagnosisofaprogressive
condition.
DoNOTuseifyouhave,
in the area being treated:
activeorsuspectedcancer
or undiagnosed pain with a
history of cancer. Stimulation
directlythroughaconrmedor
suspected malignancy should
beavoidedasitmaystimulate
growth and promote spread of
cancer cells.
Do NOT use iTouch Go
in proximity (e.g. 1 m) to a
shortwaveormicrowaveasthis
may produce instability in the
stimulator output.
Do NOT use simultaneously
with high frequency hospital
diagnostic/therapeutic
equipment. Doing so may
result in burns at the site of the
electrodes and possible damage
todevice.
Do NOT insert the wireless
probeinanyotheroriceother
thanthevagina.
Stopusingthedevice
and consult a health care
professional if you experience
anyadversereactionsfromthe
device.
Cautions
Caution should be used if you
haveableedingdisorderas
stimulation may increase blood
owtothestimulatedregion.
Caution should be used if you
havesuspectedordiagnosed
epilepsy as electrical stimulation
may affect seizure threshold.
Cautionshouldbeobserved
whenusingthedeviceatthe
same time as being connected
to monitoring equipment with
body worn electrode pads. It may
interfere with the signals being
monitored.
Caution should be used
following recent surgical
procedures. Stimulation may
disrupt the healing process.
Caution: Do not permit use by
persons unable to understand
the instructions or persons
withcognitivedisabilities,i.e.,
Alzheimer’s disease or dementia.
Cautionshouldbeobserved
when using the iTouch Go at
high strength settings. iTouch Go
has a Yellow LED light on output
socket which means the output
will exceed 10 mA (R.M.S) or 10 V
(R.M.S)averagedoveranyperiod
of 1 sec. Prolonged use at high
settings may cause muscle injury
ortissueinammation.
Not intended for use in an
oxygenrichenvironment.
Not intended for use in
conjunctionwithammable
anaestheticsorammable
agents.
The patient is an intended
operator.
Donotserviceandmaintain
thedevicewhileinuse.
Keep away from pets and
pests.
Maintenance and all repairs
should only be carried out by
an authorized agency. The
manufacturer will not be held
responsible for the results of
maintenance or repairs by
unauthorized persons.
Ifnecessary,wewillprovide
circuit diagrams, component
part lists or other information
thatwillassistauthorizedservice
personneltorepairthedevice.
Donotimmerseyourdevice
in water or place it close to
excessiveheatsuchasa
replaceorradiantheateror
sources of high humidity such
as a nebulizer or kettle as this
may cause it to cease to operate
correctly.
Keepthedeviceawayfrom
sunlight, as long-term exposure
to sunlight may affect the
silicone causing it to become
less elastic and crack.
Keepthedeviceawayfromlint
and dust, as long-term exposure
to lint or dust may affect the
sockets or cause the battery
connectortodevelopabad
contact.
Temperature&RelativeHumidity
of storage: -25°C–+60°C, 15%-
90%R.H.Temperature&Relative
Humidity of transportation:
-25°C–+60°C, 15%-90% R.H.
Therearenouserserviceable
parts. Do not attempt to open or
modify the unit. This may affect
the safe operation of the unit and
willinvalidatethewarranty.
Note: The iTouch Go has no
essential performance
characteristics.Devicefailurewill