Peak Detect mode uses the
highest and lowest of all
the samples contained in
two consecutive acquisition
intervals. Use this mode
to help detect short rapid
transitions in a waveform.
Peak Detect is not available
at faster time per division
settings.
Hi Res mode oversamples
lower-frequency signals and
calculates the average of
all the samples for each
acquisition interval. Use Hi
Res mode to provide higher-
resolution (16-bit) sample
points for lower-frequency
waveforms.
Average mode calculates and
displays the average value
for each sample point over
a user-specified number of
acquisitions. Use A
verage
mode to reduce random noise.
Trigger concepts
The oscilloscope uses the data in the waveform record to construct and display a waveform on the oscilloscope screen. However
, as the
oscilloscope is constantly acquiring samples into the waveform record, each waveform record starts at a random point of the input signal.
This means that the waveform record sample values are constantly changing, which results in a displayed waveform that is unstable or
jittering. An unstable waveform cannot be accurately measured, making this display useless except for very broad measurements (type of
signal, approximate peak to peak signal amplitude).
Figure 1: Untriggered display
What is needed is a way to set the oscilloscope to detect, or trigger on, the same signal condition or state on the input signal for each
waveform acquisition, and store the samples in the waveform record such that the same signal condition is at the same sample time
location in the waveform record.
A trigger sets when the oscilloscope detects a defined waveform condition, such as at a specified signal voltage level on the rising edge
of the signal. The oscilloscope uses that trigger condition to store the waveform samples in the waveform record so that the signal sample
point that meets the trigger condition is at the same location in the waveform record. When displayed, the waveform record for each
acquisition shows the waveform at the same trigger point, resulting in a stable waveform on the screen.
Sampling oscilloscope concepts
TBS1000C Series Oscilloscopes User Manual 51