Uncontrolled
temperature rise.
Temperature controller
connection PCB failure.
Replace.
The fan does not run
properly and/or
makes abnormal
sounds.
1. Motor failure.
2. Temperature controller
connection PCB failure.
3. Fan blade damaged.
Replace.
The temperature is
not rising.
1. Heater failure.
2. Incorrect temperature
setting on the panel [4].
1. Replace.
2. Correct the temperature
limit (increase the setting).
Temperature
overregulation.
Wrong parameter settings
in Internal Parameters.
Readjust as appropriate.
Unsatisfactory sample
incubation results.
The samples were poorly
laid out inside of the
chamber, e.g. too close to
one another or the load of
samples was too large.
Load the samples with
enough spacing between
them. The maximum load
should not be higher than 80
% of the chamber volume
capacity.
3.6. Nameplate overview