(a) Press the Setting button [16].
(b) The panel displays “SP”. The bottom [10] displays read the temperature set
value.
(c) Press the button [17].
(d) Operate the buttons [18] and [19] to set the time value read by the time
display [8] at the top. The panel displays “ST”.
(e) Press the Setting button [16] to save the changes and leave the setting menu.
(f) If no time is set, the appliance will operate non-stop.
5. The appliance will start operating according to the settings entered. The time
unit “H” or “M” begins flashing on the display [8].
6. Wait for the temperature to stabilise and load the samples into the chamber.
Keep the samples spaced enough to ensure proper air circulation.
7. The time display [8] will be counting down from the time set.
8. Once the set time counts down to zero, the time display [8] reads “End” and the
appliance beeps.
9. To repeat the operation of the appliance with the temperature and time
settings entered previously, press and hold the button [18] for 3 seconds.
10. After incubation is complete, remove the samples, turn the switch to the "O"
position, disconnect the device from the power source and open the door.
3.3.2. AUTO PID CONTROL:
* The PID routine of the appliance is designed to eliminate control errors by
reading a group of parameters settings (proportional, integral, and differential);
when the auto PID control is done, the appliance will store the parameter setting
values and revert to normal operating mode.
1. Press and hold the button [17] for a few seconds.
2. The panel displays “AT”, and the bottom display [10] reads “0”.
3. Press the button [19]. “1” appears on the display.
4. Press the Setting button [16]. “AT” starts flashing on the display.
5. During the auto PID control process, no setting can be modified. If the
overtemperature alarm is indicated, the thermal relay of the appliance will
automatically trip to cut off the supply voltage. To stop the auto PID control
process, press and hold the button [17] for 6 seconds.
3.3.3. INTERNAL PARAMETER SETTING:
Press and hold the Setting button [16] for 3 seconds. The display reads “Lc”,
prompting the user to input the access password. There are 4 passwords and each