Reverse pipetting:
For viscous liquid, do reverse pipetting
Press the operating button to the second
stop point (C) and immerse the pipette tip 2-
3mm below the liquid surface.
Slowly release the button back to the starting
position (A). Carefully withdraw the tip from
the liquid, touching it against the edge of the
container to remove any excess liquid.
Gently press the button to the first stop point
(B) to dispense the liquid. Hold the operating
button at the first stop (B) to ensure that any
liquid remaining in the tip is not included in
the delivery.
Discard the excess liquid along with the tip,
or return the remaining liquid to the original
container.
Starting position (A)
First stop (B)
Second stop (C)