Steinberg SBS-CT-123 PRO handleiding

33 pagina's
PDF beschikbaar

Handleiding

Je bekijkt pagina 6 van 33
10 11
The user must have administrator rights on a given
computer where the program and driver will be installed.
EN
2. Select „Clear All“ option. Conrm (press left button).
3. A dialogue window will appear. Conrm (press left
button).
Zero calibration:
1. Press and hold the zero calibration button (SBS-
CT-123/SBS-CT-123 PRO/SBS-CT-2000: No. 12,
model SBS-CT-123E/SBS-CT-500: No. 14) until an
acoustic signal sounds and „ZERO“ is shown on the
display.
2. Measure the same location on the included, non-
coated calibration plate a number of times. After
the initial measurement, perform the subsequent
measurements by press the probe against the plate
upon hearing a single acoustic signal;
3. To complete calibration press the calibration button
(SBS-CT-123 / SBS-CT-123 PRO/SBS-CT-2000: No.
12, model SBS-CT-123E/SBS-CT-500: No. 14). The
device will exit calibration mode.
Multi-point calibration:
1. Press the calibration button (10) (model SBS-
CT-123/SBS-CT-123 PRO/SBS-CT-2000) or enter
the menu by pressing the left button () – > Select
„Calibration“ using the up/down buttons, conrm
using left button – > select „Point Cal“, conrm using
left button > select Enable (active), conrm using
left button – > press right button until a main screen
with the sign CALIBRATION appears on the display.
(model SBS-CT-123E/SBS-CT-500);
2. Place the calibration lm on the uncoated calibration
plate (thickness indicators are shown on the lm).
3. Press the probe against the plate with the lm, raise
after the acoustic signal sounds, the reading will be
displayed.
4. Press either up or down buttons. The current
calibration point will be shown on the display, „Ptn
X“, „X“ will ash
5. Use up/down buttons to increase / decrease reading
value until the display shows the same value as
shown on the calibration lm. Use left button to
conrm.
6. Use the probe to perform subsequent measurements
for calibration lms of dierent thickness. Repeat
steps described in points 3-5.
7. To exit calibration mode, press the calibration button
(10) (model SBS-CT-123 / SBS-CT-123 PRO/SBS-
CT-2000) or enter the menu by pressing the left
button () – > Select „Calibration“ using the up/down
buttons, conrm using left button > select „Point
Cal“, conrm using left button > select Disable
(inactive), conrm using left button (model SBS-CT-
123E/SBS-CT-500);
Data upload
To upload data from the device onto a computer, install the
device driver and the EasyCoating programme found on
the CD included with the device. Then connect the device
to a computer via a USB port. The device has to be on
during data download.
Reset to factory settings
Turn the device o.
SBS-CT-123 / SBS-CT-123 PRO/SBS-CT-2000
Press and hold the zero calibration button
Use the On/O button to switch the device on whilst
pressing the zero calibration button, until information
about system rest appears on the display.
Use left button to conrm reset to factory settings.
SBS-CT-123E/SBS-CT-500
Press and hold the On/O button until information
about system rest appears on the display.
Use left button to conrm reset to factory settings.
2.4 Cleaning and maintenance
Use cleaners without corrosive substances to clean
each surface.
Do not leave the battery in the device if it will not be
used for a longer period of time.
SAFE REMOVAL OF BATTERIES AND RECHARGEABLE
BATTERIES
1.5V AAA batteries are installed in the devices.
Worn batteries must be removed from the device
proceeding analogously to their installation.
Recycle batteries
DISPOSING OF USED DEVICES
Do not dispose of this device to municipal waste systems.
Hand it over to electric and electrical device recycling and
collection point. Check the symbol on the product, user
manual and packaging. The plastics used to construct the
device can be recycled in accordance with their markings.
By recycling you are making a signicant contribution to
the protection of our environment. Contact local authorities
for information on your local recycling facility.
Rev. 03.01.2022
o CONTINOUS (continuous measurement
– function only available in model SBS-CT-123
/SBS-CT-2000: press the probe against the
tested surface, do not remove the probe)
» GROUP MODE (save Mode selection):
o DIRECT (save measurements to temporary
memory. Once the device temporary memory
is full, measurements will be overwritten; all
measurement data saved under this mode will
be lost after the device has been switched o)
o GENERAL1
o GENERAL2
o GENERAL3
o CLEAR EDDY (delete calibration data for non-
ferromagnetic materials) (relates only to the
SBS - CT- 123, SBS-CT- 123E/SBS-CT-500/
SBS-CT-2000 models)
» CLEAR ALL (delete calibration settings)
Device calibration
Deleting previous calibration settings:
1. Enter menu mode (press left button), select
„Calibration: option. Conrm (press left button).
EN
20. type of material:
F – ferromagnetic; N – not ferromagnetic
21. USB interface
2.2 Preparing for use
Battery installation
Remove the screws and cover. Insert batteries in accordance
with the markings and signs on the device. Screw the cover
back on. To change batteries, make sure the device is o.
2.3 Device use
Turning the meter on and o
Press the POWER button to turn the device on/o. The
device will also power o automatically if latent for 3
minutes (if the Auto PowerO function is activated).
First use
When switched on for the rst time, the device will enter
factory settings mode, i.e.: „Measure Mode“. Calibrate the
device before rst use (see Calibration section below).
Measurements
Do perform a measurement, place the probe against the
tested surface. Once the acoustic signal sounds, read o
the measured value.
ATTENTION! When switching the device on make sure it is
at least 5cm away from any metal items!
ATTENTION! Measurements of porous surfaces may
generate faulty readings!
ATTENTION! After switching the device on in DIR mode,
the display will be empty and when in GEN mode the last
measurement and average measurements will appear on
the display.
The user may set measurements parameters and display
them in the device menu.
DEVICE MENU:
DELETE DATA (deleting saved data):
» CURRENT DATA (current data: pressing the left
button will delete the last measurement)
» CURRENT GROUP (current group: pressing the left
button will delete the entire, most recently measured
set of data)
RECALL (last measurement)
BROWSE STAT. (measurement statistics):
» NUMBER(quantity)
» MEAN (measurement average)
» MIN (smallest measurement)
» MAX (largest measurement
OPTIONS(options):
» MEASURE MODE (measurement method selection):
o SINGLE (single measurement: press the probe
against the tested surface and then remove it
quickly and eciently)
o GENERAL4 (for all options from General1 to
General4: measurements saved to permanent
memory. Once the memory is full, the device
will continue to make measurements, but
these will not be saved or included in statistics
and „“ will be shown of the display. The
measurements will remain in the device even
after it has been switched o)
Attention: Each save mode has its own alarm and
calibration limits.
» PROBE MODE (material selection) (relates only to
the SBS - CT- 123, SBS-CT- 123E/SBS-CT-500/SBS-
CT-2000 models):
o AUTO (automatic: in this mode the meter
will automatically detect the type of tested
materials)
o MAGNETIC (ferromagnetic materials)
o EDDY CURRENT (other materials)
» UNIT SETTINGS (measurement unit selection):
o μm
o mils
o mm
» SPEED (measurement speed selection):
o NORMAL
o FAST
» LANGUAGE (language selection):
o ENGLISH
o GERMAN
o RUSSIAN
o CHINESE
» AUTO POWEROFF (activate/deactivate auto power
o):
o DISABLE (inactive)
o ENABLE (active)
LIMIT (tested material thickness limit):
» SETTINGS:
o HIGH LIMIT (tested material maximum
thickness limit)
o LOW LIMIT (tested material minimum
thickness limit)
» CLEAR (deleting pre-set limits)
CALIBRATION:
» POINT CAL (recommended calibration for error
margin of ±1-2%+1μm):
o DISABLE(inactive)
o ENABLE(active)
o CLEAR MAGNETIC (delete calibration data for
ferromagnetic materials)
o CLEAR EDDY (delete calibration data for non-
ferromagnetic materials) (relates only to the
SBS - CT- 123, SBS-CT- 123E/SBS-CT-500/
SBS-CT-2000 models)
» ZERO CAL (recommended calibration for error
margin of ±2%+1μm):
o CLEAR MAGNETIC (delete calibration data for
ferromagnetic materials)
Rev. 03.01.2022