opened, but remember not to open it until the pressure on the display
has fallen below 6Kpa!
To start a new sterilization program open and close the door again
and select the desired program.
3.3.11 Transport/Storage
For transport the unit must be completely cooled down and drained
of liquids. Close the door to prevent it from opening unintentionally.
The door should be ajar when storing the unit when it is not in use.
3.3.12 Error codes
The following error codes can appear in the display of the unit under
ERROR in the event of malfunctions:
Error
code
Meaning
E0
Problem with the temperature sensor at the chamber
surface
E1
Temperature inside the chamber is exceeded
E2
Evaporator temperature is too high
E3
Pressure and temperature cannot be matched
E4
Problem maintaining the sterilization temperature and
pressure.
E5
Door open during sterilization
E6
Working an overly long time
E7
Pressure cannot be released
3.4. Cleaning and maintenance
a) Cleaning and maintenance
Pull the mains plug and let the unit cool down completely before cleaning,
adjusting or replacing accessories and when the unit is not in use.
b) Use only non-corrosive cleaning agents for cleaning the surfaces.
c) After each cleaning, all the parts should be dried well before the unit is used
again.