Sony LMD-XH550N handleiding
Handleiding
Je bekijkt pagina 10 van 43

10
Caution
When you dispose of the unit or accessories, you
must obey the laws in the relative area or country
and the regulations in the relative hospital
regarding environmental pollution.
Warning on power connections
Use a proper power cord for your local power
supply.
1. Use the approved Power Cord (3-core mains
le
ad) / Appliance Connector / Plug with
earthing-contacts that conforms to the safety
regulations of each country if applicable.
2. Use the Power Cord (3-core mains lead) /
Ap
pliance Connector / Plug conforming to the
proper ratings (Voltage, Ampere).
If you have questions on the use of the above
P
ower Cord / Appliance Connector / Plug, please
co
nsult a qualified service personnel.
Warning on power connections for
medical use
Customers in the U.S.A. and Canada should use
the following type of power cord.
Customers in other countries or reg
ions should
use the power cord prescribed by their country or
region.
* Note: Grounding reliability can only be achieved
w
hen the equipment is connected to an
equivalent receptacle marked “Hospital Only” or
“Hospital Grade”.
Warning
The apparatus shall not be exposed to dripping or
splashing. No objects filled with liquids, such as
vases, shall be placed on the apparatus.
Warning
To prevent injury, if mounting the unit using a
mounting arm, wall fixture, or other mounting
device prepared by the customer, mount the unit
securely as described in the instruction manual
provided with the mounting device.
Check beforehand that the mounting device used
h
as sufficient strength to support the added
weight of the unit.
Check yearly that the mounting device is securely
a
ttached.
Caution
When installing, ensure the following space
around the periphery of the unit, taking
ventilation and servicing into consideration.
– Rear side: 10 cm (4 in.) or more
– Left/Right sides: 10 cm (4 in.) or more
– Bottom side: 5 cm (2 in.) or more
–Top side: 30 cm (11
7
/
8
in.) or more
Consult with qualified personnel for the following
typ
es of installation location.
– Wall mount
–Mounting arm
Caution
Do not use the device in a MR (Magnetic
Resonance) environment.
It may cause a malfunction, fire, and unwanted
mov
ement.
This equipment is not suitable for use in locations
w
here children are likely to be present.
Caution
• This unit is heavy. Make sure to unpack and
move the unit with two or more people.
• Firmly grip the bottom of this unit as shown
be
low.
U.S.A. and Canada
Plug type
HOSPITAL GRADE
*
Cord type Min. Type SJT
Min. 18 AWG
Minimum rating for
pl
ug and appliance
couplers
10 A / 125 V
Safety approval UL Listed and CSA