Sony LMD-X2710S handleiding

41 pagina's
PDF beschikbaar

Handleiding

Je bekijkt pagina 6 van 41
6
Guidance and manufacturer’s declaration – electromagnetic immunity
The LMD-X2710S is intended for use in the electromagnetic environment specified below. The customer
or
the user of the LMD-X2710S should assure that it is used in such an environment.
Immunity test
IEC 60601
te
st level
Compliance level
Electromagnetic environment –
guidance
AC input DC input
Electrostatic
di
scharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material, a
relative humidity of at least 30%
is recommended.
Electrical fast
transie
nt/burst
IEC 61000-4-4
±2 kV for power
supply lines
±2 kV for power
supply lines
Mains power quality should be
that of a typical commercial or
hospital environment.
±1 kV for input/
o
utput lines
±1 kV for input/
output lines
±1 kV for input/
output lines
Surge
IEC 61000-4-5
±1 kV line(s) to
line
(s)
±2 kV line(s) to
ea
rth
±1 kV differential
mod
e
±2 kV common
mod
e
Not applicable Mains power quality should be
tha
t of a typical commercial or
hospital environment.
Voltage dips,
short
inte
rruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
0%
U
T
(100% dip in
U
T
)
for 0.5/1 cycles
a
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25/30
cy
cles
a
(for 0.5 sec)
0%
U
T
(100% dip in
U
T
)
for 250/300
cy
cles
a
(for 5 sec)
0%
U
T
(100% dip in
U
T
)
for 0.5/1 cycles
a
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25/30
cycle
s
a
(for 0.5 sec)
0%
U
T
(100% dip in
U
T
)
for 250/300
cycle
s
a
(for 5 sec)
Not applicable Mains power quality should be
tha
t of a typical commercial or
hospital environment. If the user
of the LMD-X2710S requires
continued operation during
power mains interruptions, it is
recommended that the LMD-
X2710S be powered from an
uninterruptible power supply or a
battery.
Power
fr
equency
(50/60 Hz)
ma
gnetic field
IEC 61000-4-8
30 A/m 30 A/m 30 A/m Power frequency magnetic fields
should
be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.