GB
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in
the location in which the AC-300MD is used exceeds the applicable RF compliance level above, the AC-
300MD should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the AC-300MD.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
c The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to
6.
795 MH
z; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
Recommended separation distances between portable and m
obile RF communications equipment
and the AC-300MD
The AC-300MD is intended for use in an electromagnetic environment in which radiated RF disturbances
ar
e contro
lled. The customer or the user of the AC-300MD can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the AC-300MD as recommended below, according to the maximum output power of
the communications equipment.
Rated maximum
ou
tput p
ower of
transmitter
W
Separation distance according to frequency of transmitter
m
IEC 60601-1-2 : 2007 IEC 60601-1-2 : 2014 + A1: 2020
150 kHz to
80 MHz
d
= 1.2
P
80 MHz to
800 MHz
d
= 1.2
P
800 MHz to
2.5 GHz
d
= 2.3
P
150 kHz to
80 MHz
d
= 1.2
P
80 MHz to
2.7 GHz
d
= 2.0
P
0.01 0.12 0.12 0.23 0.12 0.20
0.1 0.38 0.38 0.73 0.38 0.63
1 1.2 1.2 2.3 1.2 2.0
10 3.8 3.8 7.3 3.8 6.3
100 12 12 23 12 20
For transmitters rated a maximum output power not listed a
bove, the recommended separation
distance
d
in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorpti
on and reflection from structures, objects and people.