Silk'n LED Neck Mask FAC02 handleiding

102 pagina's
PDF beschikbaar

Handleiding

Je bekijkt pagina 31 van 102
31
EN
Power frequency
magnetic field
(50 Hz) 1EC
61000-4-8:2001
3 A/m 3 A/m
If image distortion
occurs, the product
must be placed
farther a way from the
source of the power
frequency magnetic
field or a magnetic
shielding protection
must be installed.
The power frequency
magnetic field should
be measured to
ensure that the
installation position is
suciently low when it
is expected to be used.
Note: UT refers to the AC network voltage before the test voltage is applied.
Guidance and manufacturing statement – electromagnetic immunity – for
non-life support equipment and systems
Table 3 Electromagnetic immunity
(corresponding to Table 204 in I EC 60601-l-2/AMDl:2020 PRV)
Guidelines and manufacturer’s declaration – Electromagnetic immunity
The product is expected to be used in the electromagnetic environment specified below and
the purchaser or user of the product should ensure that it is used in this environment.
Immunity test IEC60601 test level Coincidence level Electromagnetic
environment-guide
Radio frequency
conduction
1EC 61000-4-6:
2006
Radio frequency
radiation
1EC 61000-4-3:
2010
3 V (Eective value)
150 kHz–80 MHz
3 V/m
80 MHz–2.5 GHz
3 V (Eective value)
3 V/m
Portable and mobile
radio frequency
communication
equipment should not
be used closer to any
part of the product,
including cables, than
the recommended
isolation distance.
This distance should
be calculated based
on the formula
corresponding to the
transmitter frequency.
Recommended
isolation distance
d = 1.2 √p
d = 1.2 √p
80 MHz800 MHz
d = 2.3 √p
80 MHz–2.5 GHz

Bekijk gratis de handleiding van Silk'n LED Neck Mask FAC02, stel vragen en lees de antwoorden op veelvoorkomende problemen, of gebruik onze assistent om sneller informatie in de handleiding te vinden of uitleg te krijgen over specifieke functies.

Productinformatie

MerkSilk'n
ModelLED Neck Mask FAC02
CategorieNiet gecategoriseerd
TaalNederlands
Grootte6400 MB