21 Electromagnetic emission
Table 1
Declaration – Electromagnetic emission
Emissions test Compliance
RF emissions CISPR 11 Group 1
RF emissions CISPR 11 Class B
Harmonic emissions
1EC 61000-3-2:2009
Class A
Emissions test Clause 5
Table 2
Declaration – Electromagnetic emission
Immunity test IEC 60601 test level Compliance level
Electrostatic discharge
(ESD) IEC 61000-4-2:
2008
±8 kV contact
±2 kV, ±4 kV,
±8 kV, ±15 kV air
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15kV
air
Electrical fast transient/
burst
IEC 61000-4-4:
2012
±2 kV for power supply lines
±1 kV for input/output lines
±2 kV for power supply lines
±1 kV for input/output lines
Surge
IEC 61000-4-5:
2014
±0.5 kV, ±l kV line(s) to lines
±0.5 kV, ±1 kV,
± 2 kV line(s) to earth
±0.5 kV, ±l kV line(s) to lines
±0.5 kV, ±l kV, ±2 kV line(s)
to earth
Voltage dips, short
interruptions and
voltage variations on
power supply input lines
IEC 61000-4-11:
2004
0 % UT; 0.5 cycle at 0°, 45°, 90°,
135°, 180°, 225°, 270° and 31
0 % UT; 1 cycle and 70 % cycle
UT; 25/30 cycles
Single phase: at 0°
0 % UT; 250/300 cycles
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8:
2001
30 A/m 30 A/m
Note: UT is the AC mains voltage prior to application of the best level.