5 In the ICON MENU, press the SYS-FUNC icon to access the
system app menus.
6 Access the CALIBRATION, PROCEDURE pages in succes-
sion.
7 Set in the "Expected" field, if not yet set in the calibration
setup (CALIBRATION > SETUP > Level), the nominal value
stated by the manufacturer or taken from calibration certifi-
cate, of calibrator output (e.g., 94.00) at 1KHz frequency.
000000000000
000000000000
8
Press "Play" to start the reading.
Curr.Att.Time
18:46:12
9 If the level is enough stable, press "CAL" to start the adjust-
ment procedure.
10 The "CAL" button remains inactive until the calibration proce-
dure is finished, so calibration corrections are not allowed.
11 When done, turn off the sound calibrator, slide the microphone out of the cavity.
When the calibration is completed, the date, time and correction are updated. Access to the Calibration
history panel is allowed only after the calibration is completed.
9.2 Setting the duration of a measurement
Notice!
If the measurement duration is not set, the measurement will not be interrupted unless
manually or when the memory or battery runs out.
1
In the ICON MENU, press the SETTINGS icon to access the
system and measurement configuration menus.
2
Access the SOUND MEASURE, SETUP and Measure Timer
pages in succession.
3
Select the measurement duration.
4
Press
ļƒ–
to confirm, press X to cancel operation
.
MEASURE_TIMER