XPT800 ā Sound Level Meter and Spectrum Analyzer
V1.0
Page 116 of 124
20.3 INPUT
(EDIT APP > TRIGGER > INPUT)
The trigger input can be chosen from one or more of the parameters set in the SLM screen and acti-
vated for the trigger. The parameters selectable as trigger inputs are only those for which the Ex-
ceedance on/off (bell) function has been activated in the setting panel relating to the selected SLM
parameter. Parameters selected as trigger inputs can be set to each other in OR or END logic.
20.3.1 SLM
The screen shows 12 parameters (among which there can also be coupled parameters of the type L1
minus L2), 6 related to Par1 and 6 related to Par 2. Only the parameters available for the trigger (Ex-
ceedance on) will be active; those not available (Exceedance off) will be visible but deactivated and in
grey colour; for the latter, the selection on/off checkbox cannot be activated.
The on/off checkbox can be used to activate or deactivate the parameter for its use as a trigger dur-
ing measurement.
If the logic between Par1 and Par2 is set to AND, it is possible to select one of the activated SLM pa-
rameters in Par2 for the trigger (Exceedance on). Activating the AND logic will only trigger if both Par1
and Par2 reach the values set in the exceedance panel.
20.3.2 MRK
Allows the marker name to be changed and activated by check box.
20.3.3 MIX
Trigger IN (ON/OFF): enables external trigger on Trigger I/O input