Senseca XPT800 handleiding

124 pagina's
PDF beschikbaar

Handleiding

Je bekijkt pagina 113 van 124
XPT800 Sound Level Meter and Spectrum Analyzer
V1.0
Page 113 of 124
20.1.2 Exceedance SLM mode (Exc_slm_mode)
Through the MODE parameter, it is possible to trigger on SLM acoustic parameters exceedances ac-
cording to 4 different modes described below. Four values are available: POS, NEG, IN, OUT
POS
The exceedance occurs (trigger ON) when the sound level goes from a lower to a higher value than
the Max value (threshold Max) set as the trigger activation threshold and is deactivated (trigger OFF)
when the sound level goes from a higher to a lower value than the Min value (threshold Min) set as
the trigger deactivation threshold.
Fig. 16 - Trigger mode POS
NEG
The exceedance occurs (trigger ON) when the sound level goes from a higher to a lower value than
the Min (threshold Min) value set as the trigger activation threshold and is deactivated (trigger OFF)
when the sound level goes from a lower to a higher value than the Max value (threshold Max) set as
the trigger deactivation threshold.
Fig. 17 - Trigger mode NEG

Bekijk gratis de handleiding van Senseca XPT800, stel vragen en lees de antwoorden op veelvoorkomende problemen, of gebruik onze assistent om sneller informatie in de handleiding te vinden of uitleg te krijgen over specifieke functies.

Productinformatie

MerkSenseca
ModelXPT800
CategorieNiet gecategoriseerd
TaalNederlands
Grootte27123 MB