PRO 915-2 Differential manometer / Thermometer
Page 6 of 30
2 Overview
PRO 915-2 is a professional handheld temperature and differential pressure meter with data
logging capabilities.
The meter has an internal digital differential pressure sensor. Offset and slope user correc-
tions allows compensating any possible sensor drift with time, for the best measuring accu-
racy.
Supplied with universal barbed pressure ports for connecting 4 and 6 mm inner diameter hose.
The pressure ports are screwed to the meter by a standard G 1/8 thread, allowing the user to
change them with a different type of port, if necessary, to fit his specific needs.
In combination with a Pitot tube, the meter can calculate air speed and duct flow rate. The
Pitot tube constant and the duct section are configurable.
A standard type K thermocouple probe can be connected to the temperature input. The input al-
lows taking advantage of the K-type thermocouple sensor built into some Pitot tubes.
Detection of minimum, average and maximum values. The user can reset the statistical info to
start a new statistical calculation.
Alarm thresholds can be set, to warn the user when the set values are exceeded.
The HOLD feature allows freezing the measurement on display, while the REL feature allows
showing the measurement against a value determined by the user.
The meter can be connected to a PC via the USB-C port, for data acquisition in COM interface
mode or, in mass storage mode, for viewing or downloading the files stored in the internal
memory or connecting to the ProXware application software.
Powered by 4 standard AA size alkaline batteries. For permanent operation, the instrument
can also be powered via the USB-C port by a standard 5 Vdc standard power adapter or suit-
able power bank.
The configurable auto-off feature and LCD backlight level allows for more energy saving op-
tions.
2.1 Scope of delivery
PRO 915-2 is supplied with:
Quick start guide
4 alkaline batteries, AA size
2 PVC hoses, inner Ø 4 mm, outer Ø 6 mm
USB cable
Test report
The ProXware application software is downloadable from Senseca website.

Bekijk gratis de handleiding van Senseca PRO 915-2, stel vragen en lees de antwoorden op veelvoorkomende problemen, of gebruik onze assistent om sneller informatie in de handleiding te vinden of uitleg te krijgen over specifieke functies.

Productinformatie

MerkSenseca
ModelPRO 915-2
CategorieNiet gecategoriseerd
TaalNederlands
Grootte6427 MB