PRO 915-2 Differential manometer / Thermometer
Page 10 of 30
Attention!
Use hoses with a diameter compatible with the pressure ports. The inner diameter of the
hose should not be too large to avoid pressure loss and should not be too small to avoid
mechanically forcing the ports when inserting the hose.
Caution!
Use hoses suitable for the pressure to be measured.
It is recommended that the hoses are not under pressure during connection.
It is recommended to secure the hoses to the pressure ports with hose clamps when
measuring pressures above 1 bar. GDZ-18 hose clamp is available as optional accessory.
The pressure ports are interchangeable. It is possible to unscrew the ports to replace them
with ports of a different type, provided they have G 1/8 thread. When tightening the ports,
use a suitable tool and observe the maximum torque of 2 Nm.
A K-type thermocouple sensor (e.g., the sensor built into some Pitot tubes) can be connected
to the temperature input.