Senseca PRO 245-2 handleiding
Handleiding
Je bekijkt pagina 17 van 27

PRO 241-2 / PRO 245-2 – Absolute manometers
V10
Page 17 of 27
6.4 Protected mode
The configuration of the instrument can be protected with a numeric code (PIN) that is re-
quired to access the menu or send serial commands.
By default, the protected mode is disabled. To enable the protected mode, select Settings →
PIN configuration → Activate PIN code → Yes and press OK. You are prompted to enter the
PIN, which is 0 by default. Set the PIN, if different from the default, using the down/up arrow
keys and function keys (< >), then press OK, the instrument returns to the setting screen with
protected mode enabled.
The protected mode activation is highlighted by the lck symbol in the status bar of the display.
Attention!
lck is not displayed if cor is on, because the two symbols share the same position and cor has
priority over lck.
Upon exiting the menu after activating the protected mode, the mode is immediately active. If
the menu is entered when the mode is already active (thus entering the access PIN), upon ex-
iting the menu the PIN request to re-enter the menu is reactivated 2 minutes after the last key
press (in the meantime, it is possible to re-enter the menu without entering the PIN), unless
the 2 minutes without pressing keys have already elapsed within the menu, in which case the
PIN request to re-enter the menu is reactivated immediately upon exiting the menu.
In protected mode, accessing the menu by entering the PIN does not automatically unlock se-
rial communication, and sending the PIN via the appropriate serial command [ p.19] does not
automatically unlock menu access.
Attention!
Menu items added to the list of favorite functions are not protected if accessed via the F key.
Changing the PIN:
The code can be set from 0 to 9999 (default = 0).
To change PIN, select Settings → PIN configuration → Set PIN, set PIN using down/up arrow
keys and function keys (< >), then press OK, a message appears to remind you to activate pro-
tected mode, press Help Off to close the screen.
Attention!
Changing the PIN automatically disables the protected mode if it is active. After changing
the PIN, it is necessary to reactivate the protected mode again.
Attention!
In the package of the instrument there is a sheet with a Master PIN, different for each
instrument, which allows access to the protected features in case the user forgets the PIN after
changing it.
If you lose your Master PIN sheet, you can request a copy by indicating the serial number of your
device.