PRO 241-2 / PRO 245-2 Absolute manometers
V10
Page 11 of 27
Parameter displayed numerically and graphically.
To change the unit of measurement, see the Configuration chapter [ p.13].
5.1 Hold function
The Hold function, which can be activated/deactivated by pressing the right function key,
freezes the current measurement value on the display.
The function activation is highlighted by the hld symbol in the status bar of the display.
5.2 Rel function
The Rel function displays the relative measurement against a reference value, consisting of
the measurement value at the time the function is activated.
To activate the relative measurement, press the left function key (Rel ), the measurement is
highlighted in negative, then press the left function key ( on). The symbol and the reference
value appear on display.
To disable the relative measurement, press the left function key ( off) again.
By pressing the right function key (Exit ), the instrument exits the relative measurement en-
able/disable ( on/off) mode.
The Rel command is not available:
in the screen with the statistical values;
in the screen with the graph;
while logging.