Senseca PRO 215-2 handleiding

24 pagina's
PDF beschikbaar

Handleiding

Je bekijkt pagina 16 van 24
PRO 211-x / PRO 215-x Differential manometers
Page 16 of 24
7 Measurement basics
The instrument internal pressure sensor detects the difference between the pressure applied
to the "+" input and the pressure applied to the "-" input.
Before taking the measurement, check the need to perform zeroing by leaving the pressure
inputs open (see next chapter for measurement adjustment).
By connecting only the "+" input and leaving the "" input open, the pressure difference from
the ambient pressure (relative pressure) can be measured.
Caution!
Do not exceed the maximum overpressure specified, the sensor may be damaged.
The sensor is suitable for measurement in air and non-aggressive dry gases: do not use for
measurement in corrosive gases.
Avoid performing measurements in the presence of high-frequency sources, microwaves
or strong magnetic fields.
Connect the hoses to the instrument without forcing or bending the pressure ports.
8 Measurement adjustment
The measurement can be adjusted by setting the following parameters, available in the Pres-
sure sensor Correction:
Zero point: adds an offset (P
offs
) to the measurement.
Displayed value = Measured value + P
offs
Gradient: applies a percentage correction (C
slope
%) to the measurement. Zero indicates
no correction.
Displayed value = (Measured value + P
offs
) * (1 + C
slope
% / 100)
If a measurement correction is applied by setting a value other than zero for the Zero point"
and/or "Gradient parameters, the cor symbol appears on the display.