Senseca PRO 115 handleiding

25 pagina's
PDF beschikbaar

Handleiding

Je bekijkt pagina 23 van 25
PRO 111 / PRO 115 Pt100 thermometers
Page 23 of 25
13 Technical specifications
Measurement specifications
Measuring range
-200.00…+850.00 °C / -328.00…+1562.00 °F
Resolution
0.01 °C/°F
Accuracy
±0.05 °C or ±0.06% of measured value (whichever is
higher) ± 1 digit
Linearization
According to IEC 60751
Measurement rate
2 meas./s
Long-term drift
±0.02 °C/year
Probe cable length
10 m
General specifications
Input channels
1 for 4-wire Pt100 probes
M12 connector
Storage capacity (only PRO 115)
Up to 1 million data sets, file system based (CVS files).
Each data set includes measurement and date/time
stamp.
Logging type (only PRO 115)
Automatic with manual start/stop
Logging interval (only PRO 115)
1, 2, 5, 10, 15, 30 s / 1, 2, 5, 10, 15, 30 min / 1 h
Clock
User settable RTC
Max. drift 1 min/month @ 25 °C
Display
140 x 160 dot matrix backlit LCD
Visible area 42 x 50 mm
User interface
Multilingual
PC connection
USB-C, Mass Storage Device (only PRO 115)
Power supply
4 x AA alkaline batteries
External 5 Vdc via USB C
Power consumption
20 mA typ.
Battery autonomy
> 150 h typ. continuous operation (backlight off)
Auto power off
Yes, user configurable
Operating conditions
-5…50 °C / 0…85 %RH non-condensing
Storage temperature
-25…65 °C (without batteries)
Protection degree
IP 67 (except probe connection)
IK 06
Dimensions
170 x 78 x 38 mm
Weight
340 g approx.
Housing material
ABS, TPE (side protection), Polyester (front panel)

Bekijk gratis de handleiding van Senseca PRO 115, stel vragen en lees de antwoorden op veelvoorkomende problemen, of gebruik onze assistent om sneller informatie in de handleiding te vinden of uitleg te krijgen over specifieke functies.

Productinformatie

MerkSenseca
ModelPRO 115
CategorieNiet gecategoriseerd
TaalNederlands
Grootte5710 MB