ECO 531 - Dissolved oxygen measuring device Operation and maintenance
B-H86.0.2X.DK2-4.0 Page 20 of 24
1
Unscrews the Phillips screws (A)and remove the cover.
2
Carefully replace the two Mignon AA batteries (B). Ensure that the
polarity is correct! It must be possible to insert the batteries in the
correct position without using force.
3
The O-ring (C) must be undamaged, clean and positioned at the
intended depth.
4
Fit the cover (D) on evenly. The O-ring must remain at the intended
depth!
5
Tighten the Phillips screws (A).