ECO 531 - Dissolved oxygen measuring device Measurement Basics
B-H86.0.2X.DK2-4.0 Page 13 of 24
Note
Protective gloves must be worn when commissioning, filling and servicing the sensor.
Design
α
β
χ
δ
ε
α Filling hole:
If the electrode is used at high temperatures or if it has been
stored without its protective flask for a longer period of
time, some electrolyte will be lost due to evaporation.
Please refer to Refilling description below.
β Shaft
χ Membrane head:
It will be filled with KOH electrolyte and screwed onto the
electrode shaft (no air bubbles). Damages in the membrane,
large air bubbles or air bubble rings in the membrane head
will result in erroneous measurements. This may also be the
rea-son for errors in the calibration.
The membrane head (GWOK 02) is a spare part and can be
ordered individually.
δ Platinum electrode:
If oxygen is present, it is reduced on the platinum electrode
and the sensor delivers a signal. Soiling on the platinum
electrode or between the membrane and electrode can
influence the measurement.
ε Protective flask:
The protective flask is used to moisten the membrane.
This prolongs service life of the electrode. The protective
flask contains water.
Attention
Use water only; never use potassium chloride (KCI); this is only required for storage of pH-
electrode.