ECO 531 - Dissolved oxygen measuring device Operation
B-H86.0.2X.DK2-4.0 Page 11 of 24
Parameter Values Meaning
IN,T
Factory settings
NO
Use current configuration
YES
Reset device to factory settings.
After confirming with the function-button, the
display shows: IN,T DONE
4.2 Adjustment of the temperature measuring input
The temperature input can be adjusted with the zero point correction and the gradient
correction. If an adjustment is made, you change the pre-adjusted factory settings. This is
signaled with the display text T.OF or T.SL when switching on.
1 Switch the device off.
2 Hold the down button and press the On/Off button briefly to switch on the device and open
the Adjustment menu.
3 The display shows the first parameter. Release the down button.
Parameter Values Meaning
T.OF
Zero point correction of the temperature
0.00
No zero point correction
-5.00 5.00
Zero point correction in °C. (at °F -9.00 .. 9.00)
T.SL
Gradient correction of the temperature
0.00
No gradient correction
-5.00 5.00
Gradient correction in %
Formula used by device:
Temperature = °C: Display = (measured value - T.OF) * (1 + T.SL / 100)
Temperature = °F: Display = (meas. value - 32 °F - T.OF) * (1 + T.SL / 100) + 32 °F