ECO 52x series - Handheld Conductivity Meter / EC-Meter
Error and system messages
B-H87.0.0X.DK2-4.1 Page 18 of 24
7 Error and system messages
Display
Meaning
Possible causes
Remedy
----
Range switches or
measured value
unstable
Impurities or air
bubbles
Measured value far
outside of the
measuring range
Controlling instable
Contamination or air
bubbles on cell
Measurement
outside permissible
range
Measuring cell
defect
Wait for the transient
effect of the controller
clean contamination /
bubbles
Keep measurement
within the permissible
range
Send in for repair
SENS ERRO
Sensor cable defect
Sensor or probe defect
Measuring range
exceeded/undercut
Cable breakage
Sensor or probe
defect
Measurement
outside permissible
range
Send in for repair
Send in for repair
Stay within allowable
measurement range
No display,
unclear char-
acters or no
response
when buttons
are pressed
Battery depleted
System error
Device is defective
Battery depleted
Error in the device
Replace battery
Send in for repair
BAT
Battery depleted
Battery depleted
Replace battery
ERR.1
Measuring range
exceeded
Measurement too
high
Measuring cell
defect
Stay within allowable
measurement range
Check the measuring
cell / send in for repair
ERR.2
Measuring range is
undercut
Measurement too
low
Measuring cell
defect
Stay within allowable
measurement range
Check the measuring
cell / send in for repair
SYS ERR
System error
Error in the device
Switch device on/off
Replace batteries
Send in for repair