ECO 52x series - Handheld Conductivity Meter / EC-Meter
Measurement Basics
B-H87.0.0X.DK2-4.1 Page 13 of 24
Caution!
Measuring cell must never come into contact with water-repellent materials such as oil or
silicone.
5.2.1 General Design Electrodes / Measuring cell
There are basically two different types of measuring cells: 2-pole and 4-pole measuring cells.
Control and/or evaluation take place in a similar manner; the 4-pole measuring cells can
compensate well for polarization effects and contamination to a certain degree with the
more elaborate measuring processes.
2-pole measuring cell
4-pole measuring cell
The device series is equipped with a permanently connected 2-pole measuring cell.
5.2.2 Calibration / adjustment of the measuring cell
Note
In harsh applications and due to ageing processes, the cell constant of measuring cells
changes, which leads to incorrect measured values.
Depending on the application and precision requirement, the overall precision of the display
device and measuring cell measuring chain should be checked regularly. Special testing and
calibration solutions, such as GKL 100, 101 and 102 are available for this purpose.
In normal application conditions, semi-annual testing is recommended please refer
to 4.2 “Adjustment of the measuring input“.