ECO 420 series - Compact CO2 monitor with alarm
Operation and maintenance
B-H90.0.0X.DK2-4.1 Page 18 of 26
measuring range of the sensor, the device cannot perform stability recognition. It must be
ensured that the carbon dioxide concentration at the sensor is stable before the
adjustment is started.
4 If a stable correct value is reached, the display flashes briefly, an acoustic signal sounds
and the bar indicator blinks.
5 Remove the extraction device, if applicable.
The basic sensor adjustment is finished
6.2.4 Notes on the use of the gas sampling extraction device
For adjustment with reference gas, the sensor openings of the device must be covered and a
connection piece screwed in. We offer corresponding sets for this purpose
Remove plug
Place gas orifice on gas inlet holes
Screw in the Quickconnect adapter
After the adjustment, remove the Quickconnect adapter and the gas orifice and close the
opening with the plug or the screw plug included in the adjustment set.