ECO 420 series - Compact CO2 monitor with alarm
Operation
B-H90.0.0X.DK2-4.1 Page 10 of 26
4 Operation
4.1 Opening the configuration menu
1 Press the Function key for 2 seconds to open the Configuration menu.
2 (ONF appears in the display. Release the Function key.
Parameter
Values
Meaning
INP
Measuring unit
%.
CO
2
measurement in %
PPm
CO
2
measurement in ppm
AL.
Alarm
OFF
No alarm active
ON
Alarm via text insertion, acoustic signal and flashing
of the background lighting
BEEP
Alarm via text insertion and acoustic signal
L,TE
Alarm via text insertion and flashing of the back-
ground lighting
AL.1
Pre-Alarm limit (only available if AL <> off)
0.000 .. AL.2
0 .. AL2
Alarm limit in % or ppm; a pre-alarm is triggered
when the value is exceeded
AL.2
Main-alarm limit (only available if AL <> off)
ECO 420-02
AL.1 .. 1.000
AL.1 .. 10000
Alarm limit in % or ppm, the main alarm is triggered
when the value is exceeded
ECO 420-20
AL.1 .. 3.200
AL.1 .. 19999
Alarm limit in % or ppm, the main alarm is triggered
when the value is exceeded
LCD.2
Mean value
8 H
Time weighted over 8 hours, mean value TWA
STEL
Time weighted over 15 minutes, mean value STEL
OFF
Mean value determination deactivated