ECO 410 / ECO 410T - O2-Analyser / oxygen meter
Safety
B-H86.0.31.DK2-2.1 Page 7 of 22
2.4 Intended use
The device is to be used exclusively for measuring the oxygen concentration in gas mixtures
and air. The measurement is made at the sensor opening.
The device must be calibrated regularly (in fresh air =20.95% vol. % O2) to obtain accurate
readings.
The ECO 410T is also suitable for displaying the calculated maximum operating depth (MOD)
from the measured oxygen concentration.
For the MOD display, only air and NITROX mixtures with an oxygen concentration be-tween
20.9 ... 36.0 vol. % O2 may be used as dry gases!
Danger!
For safe measurements, the function of the sensor and device must be checked regularly:
Regular calibration on ambient air (20.9 vol. % O2) in accordance with the instructions
In the case of higher oxygen values, check the function with a known gas concentration,
preferably in the area of the gas to be measured (e.g. pure oxygen, NITROX50 or similar,
"bump test")
The MOD function (calculated maximum operating depth for diving) is designed only for
use in recreational diving from 20.9 to 36.0 vol. % O2. It does not replace necessary
calculations and application of the rules for planning and performing a safe dive according
to NITROX training.
The MOD function is not designed for depth calculation of gas mixtures other than those
mentioned above (no "tech-diving", NITROX50, TRIMIX, …).
Any use of the device in this area requires appropriate training and additional safety
measures.