ECO 410 / ECO 410T - O2-Analyser / oxygen meter
Operation and maintenance
B-H86.0.31.DK2-2.1 Page 16 of 22
Note
Unnecessary unscrewing endangers the protection against moisture and should therefore
be avoided.
Read the following handling instructions before replacing batteries and follow them step by
step.
If disregarded, the device could be damaged or the protection from moisture could be
diminished.
1 Unscrews the Phillips screws (A) and remove the cover.
2 Carefully replace the two Mignon AA batteries (B).
Ensure that the polarity is correct! It must be possible to
insert the batteries in the correct position without using
force.
3 The O-ring (C) must be undamaged, clean and positioned
at the intended depth.
4 Fit the cover (D) on evenly. The O-ring must remain at
the intended depth!
5 Tighten the Phillips screws (A).