ECO 240-.. - Vacuum meter | barometer Service
B-H88.0.1X.DK2-2.0 Page 23 of 24
10 Service
10.1 Manufacturer
If you have any questions, please do not hesitate to contact us.
10.2 Calibration and adjustment service
The purpose of the calibration is to verify the precision of the measuring device by comparing
it with a traceable reference.
Both ISO calibration certificates and DAkkS calibration certificates are available from
Senseca.
Note
• The ISO standard 9001 is applied for the iso-calibration certificates.
These certificates area affordable alternative to the DAkkS calibration certificates and
provide information of the traceable reference, a list of individual values and
documentation.
• The DAkkS calibration is based on DIN EN ISO/17025, the accreditation basis is
recognized worldwide. These certificates offer high-quality calibration and consistently
high quality. The DAkkS calibration includes any necessary adjustment with the purpose
of minimizing a deviation of the measuring device.
• The device is delivered with a test report.
This confirms that the measuring device has been adjusted and tested, without making
any further statement about the accuracy.
• Only the manufacturer can check the basic settings and make corrections if necessary.
.