5.2.2 AVR 0:02 / AVR 0:05 / AVR 0: 10
Fast measurement with mean value over 2 s / 5 s / 10 s
• Mean value mode for measurement of heavily fluctuating pressures.
• In the Configuration mode, a special function AVR 0:02, AVR 0:05 or AVR 0: 10 has been
selected.
• By pressing and holding the Function key for 4 s the measurement with mean value can
be activated.
• The different mean value times of 2, 5 or 10 seconds can be selected depending on the
requirement.
• The first parameter is shown in the auxiliary display.
By pressing and holding the Function key for 2 s the special function can be deactivated.
END FVNC appears in the display.
If the tare function is active when calling up the special function AVR, it can be reset by press-
ing the function key for 4 seconds. In order to reactivate tare, the special function must be
changed in the configuration menu.