ECO 240-.. - Vacuum meter | barometer Operation
B-H88.0.1X.DK2-2.0 Page 12 of 24
4.2 Open the adjustment menu
The sensor adjustment can be adjusted with the zero point correction and the gradient cor-
rection. If an adjustment is made, you change the pre-adjusted factory settings. This is
signaled with the display text PR.OF or PR.SL when switching on.
1 Switch the device off.
2 Hold the down button and press the On/Off button to switch on the device and open the
Adjustment menu.
3 The display shows the first parameter. Release the down button.
Parameter Values Meaning
PR.OF
Zero point correction
0.00
No zero point correction
-5.00 5.00
Zero point correction [in selected display unit]
(e.g. at ECO 240-2: ± 500 hPa)
PR.SL
Gradient correction
0.00
No gradient correction of the temperature
-5.00 5.00
Gradient correction in %
Formula:
Zero point correction: Displayed value = measured value - PR.OF
Gradient correction: Display = (measured value - PR.OF) * (1 + PR.SL / 100)