ECO 240-.. - Vacuum meter | barometer Measurement Basics
B-H88.0.1X.DK2-2.0 Page 13 of 24
5 Measurement Basics
5.1 Vacuum measurements
The device is especially suitable to measure rough vacuum via the pressure port very fast.
But if the complete device is evacuated the following has to be considered:
Caution Damage possible by fast complete evacuation!
Due to the water tight construction a complete evacuation of the device may de-grade
water protection. The keypad/display screen may break loose. If the device should still be
used for this application, the following options are available:
Remove the o-ring of the battery compartment! The Instrument then is no more water
protected! Changing battery.
Specialized variants ECO 240-1-VAC!
5.2 Special functions
With the special functions that can be selected via the Configuration menu, the device can
be optimized for special measuring tasks. After it is switched on, the device starts up in stand-
ard measuring mode, the relevant special function is started by pressing and holding the
Function key for 4 s.
5.2.1 NVLL Tare function
The special function FVNC NVLL has been selected in the configuration menu.
The display can be zeroed by pressing the Function key for 4 s. If the tare function is activated,
NVLL blinks in the lower display. The tare function can be reset by pressing the Function key
again for 4 s.
Note
The tare function is independent of the zero point correction accessible via the adjustment
menu.