8 Error and system messages
Display Meaning Possible causes Remedy
---- Calculation not
possible
• Measurement data
acquisition is
running
Waiting for data
collection
No display,
unclear
characters or
no response
when buttons
are pressed
Battery depleted
System error
Product is defective
• Battery depleted
• Error in the
product
• Product is
defective
Replace battery
Send in for repair
BAT
Battery depleted • Battery depleted Replace battery
ERR.1
Measuring range
exceeded
• Measurement too
high
• Product is
defective
Stay within allowable
measurement range
Send in for repair
ERR.2
Measuring range is
undercut
• Measurement too
low
• Product is
defective
Stay within allowable
measurement range
Send in for repair
ERR.3
Display range has been
exceeded
• Incorrect display
unit
• Incorrect
resolution
• F,NE Function
active
Correct setting
Deactivate function
ERR.4
Display range has been
undercut
• Incorrect display
unit
• Incorrect
resolution
• F,NE Function
active
Correct setting
Deactivate function
SYS ERR
System error • Error in the
product
Switch product on/off
Replace batteries
Send in for repair