Senseca ECO 210-5 handleiding

24 pagina's
PDF beschikbaar

Handleiding

Je bekijkt pagina 19 van 24
ECO 210-... - Fine manometer | manometer Error and system messages
B-H88.0.2X.DK2-2.0 Page 19 from 24
8 Error and system messages
Display Meaning Possible causes Remedy
---- Calculation not
possible
Measurement data
acquisition is
running
Waiting for data
collection
No display,
unclear
characters or
no response
when buttons
are pressed
Battery depleted
System error
Product is defective
Battery depleted
Error in the
product
Product is
defective
Replace battery
Send in for repair
BAT
Battery depleted Battery depleted Replace battery
ERR.1
Measuring range
exceeded
Measurement too
high
Product is
defective
Stay within allowable
measurement range
Send in for repair
ERR.2
Measuring range is
undercut
Measurement too
low
Product is
defective
Stay within allowable
measurement range
Send in for repair
ERR.3
Display range has been
exceeded
Incorrect display
unit
Incorrect
resolution
F,NE Function
active
Correct setting
Deactivate function
ERR.4
Display range has been
undercut
Incorrect display
unit
Incorrect
resolution
F,NE Function
active
Correct setting
Deactivate function
SYS ERR
System error Error in the
product
Switch product on/off
Replace batteries
Send in for repair