Senseca ECO 141 handleiding

20 pagina's
PDF beschikbaar

Handleiding

Je bekijkt pagina 16 van 20
ECO 141 - High precision reference thermometer
Technical data
B-H86.0.07.DK2-4.1 Page 16 of 20
8 Technical data
Measuring range
0.00 .. +80.00 °C (32.00 .. +176.00 °F)
Accuracy
37 °C: ± 0.05 K ± 1 digit
0 .. +80 °C: ± 0.1 K ± 1 digit
Probe
2-wire NTC special probe, fixed connected to the device
Probe tip
Ø 1.1 x 12 mm, stainless steel
Intermediate cable
Ø <1.0 x 27 mm, teflon
Transition sleeve
Ø 4 x 45 mm
Connection cable
approx. 1.2 m, PVC
Response time t
90
<10 s (air, 2 m/s)
Measuring cycle
approx. 2 measurements per second
Display
3-line segment LCD, additional symbols, illuminated (white,
luminous duration adjustable)
Standard functions
Min/max/hold,
alarm (optical and acoustic)
Adjustment
Offset and gradient correction
Housing
Break-proof ABS housing
Protection rating
IP65 / IP67
Dimensions L*W*H
108 * 54 * 28 mm, without BNC socket or kink protection
Weight
~ 142 g incl. batteries
Nominal temperature
25 °C
Operating conditions
-20 to 50 °C; 0 to 95 %RH (temporarily condensing)
Storage temperature
-20 bis 70 °C
Current supply
2 * AA batteries (mignon)
Current requirement
approx. 0.4 mA, approx. 2 mA with backlight