ECO130 - Thermocouple Quick Response Thermometer
Error and system messages
B-H89.0.0X.DK2-2.1 Page 15 of 20
7 Error and system messages
Display
Meaning
Possible causes
Remedy
----
No suitable
measuring probe
connected
Measurement far
outside of the
measuring range
Incorrect measuring
probe
Measuring probe or
device defect
Connect a suitable
measuring probe
Measurement leaves the
permissible range
Send in for repair
No display,
unclear char-
acters or no
response
when buttons
are pressed
Battery depleted
System error
Device is defective
Battery depleted
Error in the device
Replace battery
Send in for repair
BAT
Battery depleted
Battery depleted
Replace battery
ERR.1
Measuring range
exceeded
Measurement too
high
Incorrect measuring
probe connected
Measuring probe or
device defect
Stay within allowable
measurement range
Check measuring probe
Send in for repair
ERR.2
Measuring range is
undercut
Measurement too
low
Incorrect measuring
probe connected
Measuring probe or
device defect
Stay within allowable
measurement range
Check measuring probe
Send in for repair
ERR.7
System error
Cold junction
outside operating
range or defect
Stay within allowable
operating temperature
Send in for repair
SYS ERR
System error
Error in the device
Switch device on/off
Replace batteries
Send in for repair