ECO 130.2 - Quick response 2 channel-alarm-thermometer
Error and system messages
B-H89.0.1X.DK2-2.1 Page 17 of 22
7 Error and system messages
Display
Meaning
Possible causes
Remedy
----
No suitable
measuring probe
connected
Measurement far
outside of the
measuring range
• Incorrect measuring
probe
• No 2 probes
connected for diff.
measurement
• Measuring probe or
device defect
Connect suitable
measuring probe(s)
Measurement leaves the
permissible range
Send in for repair
No display,
unclear
characters or
no response
when buttons
are pressed
Battery depleted
System error
Device is defective
• Battery depleted
• Error in the device
Replace battery
Send in for repair
BAT
Battery depleted
• Battery depleted
Replace battery
ERR.1
Measuring range
exceeded
• Measurement too
high
• Incorrect measuring
probe connected
• Measuring probe or
device defect
Stay within allowable
measurement range
Check measuring probe
Send in for repair
ERR.2
Measuring range is
undercut
• Measurement too
low
• Incorrect measuring
probe connected
• Measuring probe or
device defect
Stay within allowable
measurement range
Check measuring probe
Send in for repair
ERR.7
System error
• Cold junction
outside operating
range or defect
Stay within allowable
operating temperature
Send in for repair
ERR.1 1
value could not be
calculated
• Min. 1 measured
value invalid
Operate within the
permissible measuring
range
SYS ERR
System error
• Error in the device
Switch device on/off
Replace batteries
Send in for repair