ECO 123 - Ambient thermometer
Operation
B-H86.0.09.DK2-1.1 Page 10 of 18
Parameter
Values
Meaning
UN,T
Display unit
°(
Temperature display in °C
°F
Temperature display in °F
IN,T
Factory settings
NO
Use current configuration
YES
Reset device to factory settings.
After confirming with the function-button, the
display shows: IN,T DONE
4.2 Adjustment of the measuring input
The temperature input can be adjusted with the zero point correction and the gradient cor-
rection. If an adjustment is made, you change the pre-adjusted factory settings.
This is signaled with the display text T.OF or T.SL when switching on.
1 Switch the device off.
2 Hold the down button and press the On/Off button briefly to switch on the device and
open the Adjustment menu.
3 The display shows the first parameter. Release the down button.
Parameter
Values
Meaning
T.OF
Zero point correction
0.00
No zero point correction
-5.00 5.00
Zero point correction in °C. (at °F -9.00 .. 9.00)
T.SL
Gradient correction of the temperature
0.00
No gradient correction of the temperature
-5.00 5.00
Gradient correction in %
Formula used by device:
Temperature = °C: Display = (measured value - T.OF) * (1 + T.SL / 100)
Temperature = °F: Display = (meas. value - 32 °F - T.OF) * (1 + T.SL / 100) + 32 °F