9 Technical data
Measuring range
-70.0 .. +250.0 °C (-94.0 .. +482.0 °F)
Accuracy
-20 .. +100 °C: ± 0.1 K ± 1 Digit
-70 .. +250 °C: ± 0.2 % of measured value ± 2 digits
Probe
ECO 121-H-D
Immersion probe Ø 3 x 150 mm, V4A, large handle
(temperature resistance up to 120°C), silicone cable
ECO 121-H-I1.5
Extra-thin and fast insertion probe Ø 1,5 x 100 mm, V4A,
compact silicone handle (temperature resistance permanent
230°C / short time 250°C), silicone cable
ECO 121-H-I3
Robust insertion probe Ø 3 x 100 mm, V4A, large handle
(temperature resistance up to 120°C), silicone cable
ECO 121-H-K
Robust PTFE- insertion probe Ø 3 x 100 mm, V4A, large
PTFE handle, PTFE cable
(completely temperature-resistant up to 250°C)
Response
time t
90
water (0,4
m/s)
…-D
<3 s
…-I3, …-K
<3 s
…-I1.5
<2 s
Measuring cycle
approx. 2 measurements per second
Display
3-line segment LCD, additional symbols, illuminated (white,
luminous duration adjustable)
Standard functions
Min/max/hold, auto hold, timer function
alarm (optical and acoustic)
Adjustment
Offset and gradient correction
Housing
Break-proof ABS housing, polyester protection screen /
keypad
Protection rating
IP65 / IP67
Dimensions L*W*H
108 * 54 * 28 mm, without kink protection
Weight
~ 180 g incl. batteries and probe (ECO 121-H-D)
~ 150 g incl. batteries and probe (ECO 121-H-I1.5)
~ 175 g incl. batteries and probe (ECO 121-H-I3)
~ 190 g incl. batteries and probe (ECO 121-H-K)
Nominal temperature
25 °C