ECO 120 / ECO 121 - Alarm thermometer
Operation
B-H86.0.0X.DK2-4.1 Page 12 of 22
Parameter
Values
Meaning
T.SL
Gradient correction of the temperature
0.00
No gradient correction of the temperature
-5.00 … 5.00
Gradient correction in %
Formula used by device:
Temperature = °C: Display = (measured value - T.OF) * (1 + T.SL / 100)
Temperature = °F: Display = (meas. value - 32 °F - T.OF) * (1 + T.SL / 100) + 32 °F
Adjustment example:
Temperature unit= °C, the adjustment is carried out in 2 separate steps at 0 °C (e.g. ice wa-
ter) and at a reference temperature (e.g. clinical thermometer with water bath 37 °C).
First set the values of t.oF and t.SL in the adjustment menu to 0
Zero point:
Set the temperature probe to 0 °C and let it adjust.
Start the adjustment menu and enter the display value at 0 °C for t.oF
After leaving the menu, the device should now display 0.0 °C.
Slope:
Set the temperature sensor to the reference temperature and let it
adjust.
Calculate slope correction:
Start adjustment menu and enter the calculated value at t.SL
Exit menu, the device should now display the reference temperature.