2. About broncolor Radio System
●
NOTE
● With broncolor radio system, individual ash triggering is not available for each Lamp
Address. All ashes in the same Studio Address will re even if a particular Lamp Address is
selected. For measuring the ash individually, turn off the power of ash units not in use.
● When the measured value display area is touched, both ash waveform graph and measured
value are displayed. When it is touched again, the display returns to the previous screen.
● Measure the ash light characteristics in a darkroom without ambient light.
● To set ISO sensitivity and shutter speed, press the [Flash Duration Analysis Radio
Triggering Mode] Icon ( ) to return to the Measuring Screen.
● The last selected Lamp Address in either Flash Power Control Screen or Radio Studio/
Lamp Screen is activated in Measuring Screen.
● For radio frequencies, refer to “5. Radio Studio Address Frequencies”. (
P36)
[Measuring Screen]
[Flash Duration Analysis
Mode Flash Waveform
Graph Screen]
Touching the
Measured
value area
changes the
display.
* The graph screen cannot be used to make measurements.